Alghamdi Ali D
King Fahad Hospital, Saudi Arabia
Title: Infl uence of dimensions on the primary stability and removal torque of short dental implants.
Biography
Biography: Alghamdi Ali D
Abstract
Background: Reduced vertical bone level in the implantation area is oft en considered one of the limiting factors before implant insertion. Inserting implants of reduced length might be useful in order to avoid vertical bone augmentation prior to implantation. Th e use of short implants in compromised sites are considered an alternative procedure to avoid extensive surgical procedures such as sinus lifting and grafting. Methods: 40 short dental implants (4.8 and 6.2 mm diameter with 5 and 7 mm diameter). fixtures were installed on 20 bovine rib blocks. Th e primary stability of the implant was measured by the resonance frequency using an Osstel ® device. Th e removal torque values (RTV) of the implants was assessed using a Digital torque gauge instrument. Results: Both 4.8 and 6.2 mm wider implants showed marginal increase in resonance frequency and removal torque values with 5 and 7 mm implants. However when the comparison was done between the two implants with different diameter a significantly higher primary stability was observed with 6.2 mm diameter implants. Conclusion: From the observations of the study it can be concluded that short implants were able to achieve desirable primary stability. Th e primary stability substantially improved with short implants with wider diameter.